Biomechanics of the Injured SIJ: Results from an In Vitro Study



Fig. 4.1
Specimen mounted on six degree-of-freedom spine tester with rigid body markers for motion tracking





Range of Motion


Plexiglass markers, each having three infrared light-emitting diodes, were secured rigidly to L3, L4, L5, sacrum, left iliac crest, and right iliac crest via bone screws to track motion using the Optotrak Certus (NDI, Inc., Waterloo, Canada) motion analysis system. The location of the markers (denoting a rigid body) is aligned approximately sagittal along the curvature of the spine. The Optotrak Certus software superimposes the coordinate systems of two adjacent vertebral bodies in order to inferentially determine the relative Eulerian rotations in each of the three planes.


Surgical Reconstruction Groups


All specimens were tested intact with no damage to ligaments or bones. The REVERE® Stabilization System (Globus Medical Inc., Audubon, PA) was used for the 5.5 mm-diameter rigid titanium rod construct. Each of the seven specimens were tested in the following order: (1) intact; (2) left side sacrotuberous ligament (cephalad portion) injury (L-STL cut); (3) left side dorsal sacroiliac and interosseous ligaments injury (L-DSIL + L-IL cut), with the cumulative injury resulting in a transection of the entire left posterior ligament complex (L-PL complex cut); and (4) L5–S1 rigid instrumentation (L5–S1 rigid) placed adjacent to the posterior ligament injured SIJ. L-STL cut construct was performed by carefully dissecting the cephalad portion of the sacrotuberous ligament at a point just below the posterior superior iliac spine. Rongeur and a scalpel were used to completely transect the ligament (Fig. 4.2).

A317803_1_En_4_Fig2_HTML.jpg


Fig. 4.2
Specimen with transected posterior sacrotuberous ligament (cephalad portion)

The L-PL complex cut construct was created by inserting the scalpel into the entire junction of the posterior iliac crest and sacrum from its most cephalad to its most caudal borders transecting all of dorsal sacroiliac and interosseous ligaments down to the joint (Fig. 4.3).

A317803_1_En_4_Fig3_HTML.jpg


Fig. 4.3
Specimen, with previous transection of the posterior sacrotuberous ligament, now with the additional transection of the dorsal sacroiliac ligament and the interosseous ligament. This constitutes a transection of the entire posterior supporting ligament structure for the SIJ



Statistical Analysis


The data points include one range of motion measurement per specimen, under uniform stresses, for three different directions. These measurements were taken at each of the ten stages mentioned above, including intact, simulated injuries and stabilization constructs. For raw data, see Appendix B.

To test the hypotheses, analysis involved paired comparisons of range of motion measurements between each of the first four successive measurements. The Shapiro–Wilks normality test was applied to the measurements for each stage to check assumptions of the paired t-test. While more than half of the individual samples were found to diverge significantly from normality, it is acknowledged that the power of any such test is rather limited for such a small sample size. Therefore, the Wilcoxon sign test was also implemented. Note the agreement of significance (at α = 0.05) between the two tests in Table 4.1.


Table 4.1
Comparisons of range of motion measurements



















































































Flexion–extension

Intact → STL cut

STL cut → PLC cut

PLC cut → L5–S1 rods

Change in ROM tested

Increase

Increase

Increase

t-stat (p-val)

0.0258

0.0316

0.0082

W-sgn (p-val)

0.0391

0.0078

0.0078

Significance

Yes

Yes

Yes

Lateral bend

Intact → STL cut

STL cut → PLC cut

PLC cut → L5–S1 rods

Change in ROM tested

Increase

Increase

Increase

t-stat (p-val)

0.6321

0.0518

0.7157

W-sgn (p-val)

0.7693

0.0574

0.7081

Significance

No

No

No

Axial rotation

Intact → STL cut

STL cut → PLC cut

PLC cut → L5–S1 rods

Change in ROM tested

Increase

Increase

Increase

t-stat (p-val)

0.5539

0.0283

0.1849

W-sgn (p-val)

0.6238

0.018

0.3057

Significance

No

Yes

No

Small sample sizes leave a question of power in the ability to detect differences. Using the average sample standard deviations over three directions, e.g.,
Oct 22, 2016 | Posted by in ORTHOPEDIC | Comments Off on Biomechanics of the Injured SIJ: Results from an In Vitro Study