Total Shoulder Arthroplasty: Epidemiology and Causes


Fig. 7.1

In these images, the picture on the right shows an early glenoid component loosening, 1 year after the index procedure (picture on the left). A guided aspiration was obtained, and cultures were positive for C. acnes infection



More “classic” clinical findings suggesting infection may be encountered in the setting of a more aggressive organism, such as Staphylococcus or Streptococcus spp. In these instances, bone osteolysis and implant loosening, swelling, erythema, and increased blood infection markers may be present [12, 21, 25].


Diagnosis of infection can often be difficult, with pain and limited range of motion being the most common clinical complaints [26]. Good-quality radiographs can help rule out conditions that may mimic or coexist with an infected shoulder arthroplasty, such as post-arthroplasty rotator cuff failure. It is common practice to obtain a baseline laboratory analysis with white blood cells (WBC) (percentage polymorphonuclear cells), erythrocyte sedimentation rate (ESR), and C-reactive protein (CRP). Computed tomography (CT) scans can be useful to detect osteolysis and to assess remaining bone stock. Ultrasonography and magnetic resonance imaging (MRI) with metal subtraction protocols can determine the presence of local abscesses, effusion, or osteomyelitis. Scintigraphy can detect inflammation but may not be useful in low-grade infections [27, 28].


Synovial fluid analysis from an aspiration or at the time of revision surgery should include cell count, gram stain, cultures for aerobes, anaerobes, fungi, and mycobacteria and should be held for up to 4 weeks [24]. Unfortunately, a negative culture or gram stain does not always rule out infection. Intraoperatively, at least five biopsy samples should be sent for gram stain and frozen section [29]. Interestingly, increased body mass index, diabetes severity, and asymptomatic bacteriuria or abnormal urinalysis have not been associated with increased rates of infection [12, 3032].


7.3.2 Instability


7.3.2.1 Instability After Anatomic Total Shoulder Arthroplasty (TSA)


Instability after anatomic total shoulder arthroplasty is a relatively common complication, with a reported prevalence ranging from 1% to 3% [11, 33]. It can occur secondary to insufficient bone stock, inadequate soft tissue balance, component malalignment, or loosening.


Severe primary osteoarthritis, as well as post-capsulorrhaphy arthritis, can lead to excessive acquired retroversion of the native glenoid. Anterior wear is more uncommon, but it can be found in patients with chronic anterior glenohumeral dislocations, glenoid fractures, or rheumatoid arthritis. Failure to identify and correct this deformity can result in glenoid component malalignment and either posterior or anterior instability. Humeral component malpositioning is usually less critical, but it can also play a role in instability.


Diagnosis can be difficult, and a careful physical examination is paramount. In some patients, dislocation of the glenohumeral joint can be obvious radiographically, but in the setting of subluxation, findings will be more subtle. Excessive translation of the humeral head or a positive load-and-shift test can help the examiner in the diagnosis of these cases [34, 35].


Anterior instability after anatomic total shoulder arthroplasty has been reported in 0.9% of the patients [11] and has been associated with subscapularis failure, retroversion of the humeral component of less than 20° [36], anterior glenoid deficiency, and anterior deltoid dysfunction [11]. Of these causes, it is thought that subscapularis dysfunction plays a major role. In these patients, a positive lift-off test and/or belly press test can be found [37].


Management of the subscapularis during the initial surgery remains controversial, as some authors report improved outcomes after a lesser tuberosity osteotomy (LTO) versus a tenotomy or peel technique [38]. This clinical finding has been supported by biomechanical analyses [39, 40]. However, the peel technique or tenotomy of the subscapularis avoids the potential complication of LTO nonunion [41]. To date, there is insufficient high-level clinical evidence to strongly support one technique over the others.


In addition to technique, overstuffing the joint with an excessively large humeral head and medialization of the tendon insertion may lead to failed subscapularis failure. Excessively early mobilization, aggressive physical therapy, or postoperative trauma can also disrupt the subscapularis tendon repair.


Posterior instability after TSA occurs with a similar frequency as anterior instability (1%) [11] and has been associated with soft tissue imbalance. While posterior rotator cuff dysfunction and capsular laxity have been most commonly implicated, component malalignment and posterior bone loss can also play a role [34, 42]. Glenoid retroversion over 20° and humeral component in more than 45° of retroversion have been described as potential causes of posterior instability [36]. Sanchez-Sotelo and colleagues recommended that surgeons pay close attention to the humeral neck cut angle and the subscapularis tendon repair and address any posterior glenoid bone loss to minimize the potential for this complication. In addition, posttraumatic osteoarthritis or preoperative humeral subluxation should be carefully evaluated [34].


Rotator cuff failure is one of the most common complications after anatomic total shoulder replacement. A recent analysis of complications reported to the US Food and Drug Administration (FDA) demonstrated that among all the complications found after 1673 anatomic total shoulder replacements, posterior-superior rotator cuff and subscapularis failure were second only to glenoid component failure, representing 15.4% of all the complications [9]. Rotator cuff failure allows the humeral head to migrate proximally, leading to superior instability (Fig. 7.2). Reported in up to 3% of cases [11], superior instability may be the single most common direction of instability following anatomic shoulder arthroplasty.

../images/479782_1_En_7_Chapter/479782_1_En_7_Fig2_HTML.png

Fig. 7.2

Rotator cuff failure can lead to superior instability. The radiograph on the left demonstrates proximal migration of the humerus, which led to glenoid component fixation failure through the so-called rocking horse mechanism. This patient eventually underwent revision surgery to a reverse shoulder arthroplasty (right)


The rotator cuff can be compromised during the index procedure, specifically if an aggressive humeral resection is performed or if the cut is placed in too much retroversion [43]. Postoperative rotator cuff failure can also occur, with reported rates from 1.3% to 5.8% [11, 37]. Several factors have been found to affect superior instability: fatty infiltration of the infraspinatus, rotator cuff tear size, coracoacromial arch insufficiency, anterior deltoid dysfunction, humeral head overstuffing and malpositioning, and tuberosity nonunion in the setting of fracture [44, 45].


Inferior instability often occurs when the humeral length is not restored and deltoid tensioning is therefore not achieved. This has been reported to be more common after four-part proximal humerus fractures, where the stem can be accidentally seated too low due to a loss of anatomic references. Warren recommends inferior distraction of the humerus to detect this issue intraoperatively. When this maneuver is performed, the head should ideally remain within the upper one-third of the glenoid. Inferior instability may also occur in a setting of an axillary nerve palsy or rotator interval insufficiency in which the dynamic stabilizers are inadequate to hold the glenohumeral joint reduced [42].


7.3.2.2 Instability After Revision Total Shoulder Arthroplasty (RTSA)


Trappey et al. reported an instability rate after RTSA of 5% following primary cases and 8% following revision arthroplasty [46]. The mechanism of dislocation is typically adduction and internal rotation and most commonly occurs within the first 3 months following surgery. Up to 50% of these will have good outcomes with conservative treatment after successful closed reduction. Late dislocations that occur over 3 months after the index procedure often require surgical treatment [47].


Abdelfattah et al. proposed a classification system for instability after reverse total shoulder arthroplasty. They described three main categories: loss of compression, loss of containment, and impingement.


They further divided loss of compression into undersized implants, loss of deltoid contour, humeral height loss, subscapularis deficiency, acromial/scapular fracture, and deltoid dysfunction (Fig. 7.3).

../images/479782_1_En_7_Chapter/479782_1_En_7_Fig3_HTML.jpg

Fig. 7.3

This patient sustained an early dislocation after a reverse total shoulder arthroplasty (left). He underwent a closed reduction, but examination under anesthesia revealed instability of the implant (center). Therefore, revision to a larger glenosphere and a retentive polyethylene was warranted (right)


Loss of containment can be subclassified into alteration of depth/radius ratio of the humerosocket and mechanical failure (such as glenosphere-baseplate dissociation, stem fracture, or humerosocket dissociation at the trunnion).


Impingement can occur in a setting of a large body habitus, with the axillary soft tissue creating a levering-out effect with traction from the weight of the arm. Furthermore, soft tissue or bony impingement can occur in a fracture setting with unreduced retained tuberosities, malunion, or heterotopic ossification; prosthetic malalignment may play a role if the humeral component prematurely contacts the glenoid neck in adduction [48]. This can be modified by changing the glenosphere size, the baseplate placement, offset or tilt or the neck-shaft angle, and version of the humeral component [49].


Trappey and colleagues also found that patients with an irreparable subscapularis had a higher rate of instability [46]. A meta-analysis by Matthewson et al. concluded that subscapularis repair decreases the rate of instability, and in those cases when it cannot be repaired, a lateralized center of rotation results in significantly lower dislocation [50]. Owing to the preservation of the subscapularis tendon insertion, a superior subscapularis-sparing approach may lower the risk of dislocation, with reported rates of instability as low as 0%. However, glenoid exposure and baseplate placement using this approach may be significantly more challenging [51]. Subscapularis involvement in RTSA instability remains controversial in the existing literature, as similar clinical results with or without subscapularis repair have been reported [52].


7.3.3 Component Loosening


7.3.3.1 Anatomic Total Shoulder Arthroplasty Loosening


Prosthetic loosening has been reported to represent 12.4–39% of the complications after anatomic total shoulder arthroplasty [11]. Radiolucencies, calcar resorption, or scapular notchings are common findings after anatomic and reverse shoulder replacement, but not all of them may be clinically relevant. In the presence of pain or gross implant migration, however, further investigation is warranted.


Glenoid component loosening occurs more frequently than aseptic humeral component loosening, representing over 80% of fixation failures [11]. Positive radiographic findings of lucencies about the component vary from 12% to 94% in the literature, but these do not necessarily correlate with clinical findings. In this regard, surgical technique must be meticulous, as it has been suggested that the presence of lucent lines and further frank loosening may be related to the presence of cement on the backside of the glenoid component. This may indicate suboptimal bone preparation of the native glenoid and/or suboptimal seating of the component [53, 54].


Loosening can occur due to uneven force distribution in the setting of glenohumeral instability (the so-called rocking horse mechanism) [55] due to proximal migration of the humeral head in the setting of rotator cuff failure or due to infection, lack of bone stock, or poor bone fixation. Shoulder biomechanics may also play a role. Compared to other joints, the humeral head appears to have larger “play in the socket,” which may explain the faster polyethylene wear that has been found in explanted shoulder liners when compared to equivalent hip inserts [56].


Papadonikolakis found an asymptomatic radiolucency rate of 7.3% per year and symptomatic loosening of 1.2% per year, with more asymptomatic lucencies found in keeled versus pegged implants [57]. Biconcavity of the native glenoid and increased glenoid retroversion may also lead to increased component loosening. Walch et al. found a 21% loosening rate in biconcave glenoids and a 44% complication rate associated to retroversion greater than 27° [58].


Others have found that metal-backed glenoid implants have a revision rate up to three times higher than all-polyethylene components [57]. The Australian registry demonstrated an increased revision rate in both fixed and modular metal-backed glenoid components. They reported a significantly higher revision rate of non-cross-linked vs. cross-linked glenoid components with a hazard ratio of 2.38, but they found no differences in the revision rate between cemented versus hybrid glenoid components in total shoulder arthroplasties [4].


Stem aseptic loosening is much less common than glenoid failure, accounting for 7% of the complications after TSA [11]. In defining stem loosening in non-cemented stems, Sperling described eight radiographic zones around the humeral stem and concluded that a humeral component was “at risk” if a lucent line 2 mm or greater was found in at least three zones [59]. Sanchez-Sotelo used the same parameters to successfully evaluate radiographic loosening in cemented stems [60]. Changes at the bone-implant interface on the humeral side in the presence of a glenoid component have raised concerns about osteolysis and symptomatic loosening in the setting of polyethylene particle debris [45, 60].


7.3.3.2 Reverse Total Shoulder Arthroplasty Loosening


Boileau reported that among all the causes that led to revision surgery after a failed RTSA, 21% were due to humeral side complications. It was the second most common cause of revision after instability. He found that humeral loosening was often related to biological causes (polyethylene wear and metallic debris), in addition to mechanical causes (rotational forces) [47]. Radiographic loosening is rare, with a reported prevalence of less than 1% [61], but proximal humerus bone loss in a proximal humerus fracture setting, for instance, can decrease mechanical strength of the humeral stem leading to an increased risk of humeral-sided failure [62].


Glenoid component loosening is uncommon in the setting of RTSA and can be minimized by careful surgical technique [63]. Avoidance of superior tilt, placement of the baseplate at the most inferior aspect of the glenoid, and achievement of adequate primary stability that allows bone ingrowth are paramount [47, 63].


The influence of scapular notching on glenoid component loosening after reverse shoulder arthroplasty remains controversial, as some series report increased loosening rates related to scapular notching (Fig. 7.4) [14, 64, 65], while others report no association [15, 66, 67]. The use of a superior approach has been reported to increase prevalence of scapular notching [67], which suggests that this approach may indirectly increase the risk of loosening. Lateral and inferior offset of the glenosphere, on the other hand, may minimize radiographic loosening, though some lateralized designs have been reported to potentially lead to a higher rate of component dissociation [68].

../images/479782_1_En_7_Chapter/479782_1_En_7_Fig4_HTML.jpg